2 • Features
Doc. 065291-01 3/09
23
2.2
Flow Schematics
•
Figure 2-10
illustrates the liquid flow path when the components required for
producing KOH, LiOH, MSA, or NaOH eluent are installed. The required
components include the corresponding type of EluGen cartridge and a CR-
TC.
•
Figure 2-11
illustrates the liquid flow path when the components required for
producing carbonate/bicarbonate eluent are installed. The required
components include an EGC II K
2
CO
3
EluGen Cartridge, an EPM
Electrolytic pH Modifier, and an EGC CO
3
Mixer.
•
Figure 2-12
illustrates the liquid flow path when the components required for
producing carbonate eluent are installed. The required components include an
EGC II K
2
CO
3
EluGen Cartridge and an EGC CO
3
Mixer.