Oritron Dionex Ion Chromatography System ICS-2100 User Manual
Page 176
ICS-2100 Ion Chromatography System
164
Doc. 065291-01 3/09
Conditioning the EluGen Cartridge
Always condition the EluGen cartridge for 30 minutes before initial use.
1. Make sure the eluent reservoir is filled with ASTM filtered, Type I
(18-megohm) deionized water.
2. Verify that the suppressor and the auxiliary electrolytic device (if
installed) are off.
3. Verify that the
Automatically turn on with pump
option is not
selected on the ICS-2100 touch screen
SUPPRESSOR
). Disabling this option lets you turn on the pump, without
automatically turning on the suppressor.
4. From the ICS-2100 Control panel in Chromeleon or the touch screen
HOME
page, set the pump flow rate to 1.00 mL/min.
5. Turn on the pump.
6. Select the following settings from the Chromeleon Control panel or
the touch screen
ELUENT GENERATOR
page:
•
Set the
EGC-1
concentration to 9 mM.
•
Set the
EGC-2
concentration to 1 mM.
7. Turn on the EGC-1 and EGC-2 currents.
8. Run at these settings for 30 minutes.
9. Turn off the EGC-1 and EGC-2 currents.
10. Turn off the pump.
11. Remove the backpressure coil.
12. Reconnect the
TO INJ VALVE IN - P/ELUENT OUT
line to the mixer
inlet.
13. Reconnect the
TO SRS/AES REGEN OUT
line to the suppressor
REGEN
OUT
port.
To avoid damaging the suppressor (and auxiliary electrolytic device, if
installed), always turn them off before conditioning the EluGen
cartridge. The pump flow is on during conditioning; however, no flow
reaches the suppressor or auxiliary electrolytic device.