Suppressor Page
Use the
SUPPRESSOR
page to specify the type of suppressor installed in the ICS-
2100, turn the suppressor on and off, and select the amount of current supplied to
the suppressor. To open this page, touch the suppressor button on the
HOME
page,
or select
SUPPRESSOR
from the menu of pages.
Suppressor Controls
•
The suppressor type is typically selected at initial installation and does not
need to be changed. However, if a different type of suppressor is installed
later (or if the suppressor is removed from the system), touch the
Type
field
and select the new suppressor type (
None/MMS
,
SRS
, or
AES
) from the list.
•
To set the current, touch the
Current
field and enter the desired setting (in
milliamperes) on the number pad.
The appropriate suppressor current setting depends on several variables: the
suppressor type, the column, the eluent, and the flow rate. Refer to the
suppressor manual for the recommended current setting for your application.
The suppressor manuals are included on the Dionex Reference Library CD-
ROM (P/N 053891).
Figure B-18. Suppressor Page