2 removing the cr-tc and reinstalling it, In the new egc holder, Section 5.23.2 – Oritron Dionex Ion Chromatography System ICS-2100 User Manual
Page 200
ICS-2100 Ion Chromatography System
188
Doc. 065291-01 3/09
5.23.2 Removing the CR-TC and Reinstalling it in the New EGC
Holder
1. Turn the holder upside down to view the CR-TC.
2. Grasp the top of the CR-
TC and lift up to
disconnect it from the
holder.
3. Disconnect the liquid
lines from the four ports
on the CR-TC (see
Figure 5-59
).
Figure 5-76. Disconnecting CR-TC Liquid Lines
4. Turn the new holder
upside down (see
Figure 5-77
) and
connect the liquid lines
to the CR-TC.
NOTE To help retain the
fittings and ferrules
on the lines, remove
the coupler from
each line just before
connecting the line
to a port.
Figure 5-77. EGC Holder (upside down)
Disconnect liquid
lines from 4 ports
Couplers
Metal stud for mounting the CR-TC