B • Touch Screen Operation
Doc. 065291-01 3/09
229
below 30%. At 0%, an error message informs you that the ion count is
depleted and the cartridge must be replaced before continuing
operation.
NOTE When the EGC-2 power supply is configured as
an auxiliary power supply, the EGC-2 Life
Remaining field is blank.
•
Normalized Life Used
(displayed when an EPM is installed)
indicates how much the EPM has been used. The value starts at 0 and
counts up with use. A count of 1 means the EPM usage is
approximately equal to the lifetime of one K
2
CO
3
cartridge.
B.8.5
Controlling the CR-TC Power
•
To manually control the CR-TC power, touch the
CR-TC Off
and
On
buttons on the
ELUENT GENERATOR
page (see
Figure B-17
).
•
To automatically control the CR-TC power simultaneously with the
pump flow, select
Auto turn on with pump
.
NOTE The CR-TC power always turns off when the pump
is off, regardless of whether the “Automatically turn
on with pump” option is selected.
Figure B-17. Eluent Generator Page: CR-TC Controls