15 replacing the leak sensor – Oritron Dionex Ion Chromatography System ICS-2100 User Manual
Page 149
5 • Service
Doc. 065291-01 3/09
137
12. Align the new valve on the component panel with the liquid ports facing up
and attach it with the two screws removed in
Step 5
. Reconnect the liquid
lines.
13. Turn on the power to the ICS-2100.
14. Prime the pump (see
5.15 Replacing the Leak Sensor
1. Turn off the power to the ICS-
2100.
2. Open the ICS-2100 front door.
3. Loosen the screw on the front of
the leak sensor (see
). Note: The screw
remains attached to the sensor.
Figure 5-29. Leak Sensor
4. Remove the leak sensor from
the component panel and pull
the cable out of the opening in
the panel.
5. The cable attached to the leak
sensor is connected to a
matching cable inside the ICS-
2100. Continue pulling the cable
until the connectors for the two
cables are outside of the panel
(see
6. Disconnect the two cables.
Figure 5-30. Leak Sensor Cable
Leak
Sensor
Loosen
Screw
Disconnect
Cables