B.7.2, Setting degas operating parameters (optional) – Oritron Dionex Ion Chromatography System ICS-2100 User Manual
Page 236
ICS-2100 Ion Chromatography System
224
Doc. 065291-01 3/09
•
To change the low pressure limit, touch the
LOW
field and enter the
desired limit on the number pad.
To select the pressure unit for all touch screen pages that display pressure
values, touch the pressure unit field and select the desired unit on the list
(
PSI
,
bar
, or
MPa
) (see
Figure B-14
).
B.7.2
Setting Degas Operating Parameters (Optional)
By default, the ICS-2100 monitors the degas pressure reading and turns
the degas pump on and off as required. To change the degas operating
option:
1. Select
PUMP
from the menu of pages.
2. Under
Degas Settings
, touch the options drop-down list. A list of
available options appears (see
NOTE Degas options can also be selected from the
Chromeleon Server Configuration program
(see
Figure B-14. Pump Page: Selecting the Pressure Unit
Figure B-15. Pump Page: Selecting Degas Options
Select the
pressure unit
Select the
degas option