Oritron Dionex Ion Chromatography System ICS-2100 User Manual
Page 261
C • TTL and Relay Control
Doc. 065291-01 3/09
249
Connect the Analog Output
1. Locate the analog cable for the third-party analog-to-digital converter and
controller.
2. Connect the analog input connector on the third-party controller to the
ANALOG OUTPUT
connector on the ICS-2100 rear panel (see
Figure C-6
).
Configure the ICS-2100 Analog Output
Go to the ICS-2100
DETECTOR
touch screen page and select the settings shown in
Figure C-9
:
Create an AS Method and Schedule
Figure C-10
is an example of AS method timed events that can be used with the
TTL/Relay connections shown in
Figure C-6
.
Figure C-9. Detector Page
Figure C-10. Example AS Method for Stand-Alone Operation
Help Prompt
METHOD:
TIMED EVENTS
1
1
1
TIME
INIT LOAD
INJECT
0.00
VALVE
TTL1 TTL2 RLY1 RLY2
1
CSV
0
0
1
0.50