5 • service, Hydrating the cr-tc, Figure 5-62. cr-tc plumbing for hydration – Oritron Dionex Ion Chromatography System ICS-2100 User Manual
Page 179: Egc vent
5 • Service
Doc. 065291-01 3/09
167
15. Reinstall the EluGen cartridge in the holder.
16. Remove the cap on the luer adapter by twisting. Save the cap.
17. Twist the luer lock and
EGC VENT
line into the luer adapter.
18. Do not connect the EluGen cartridge cable until after hydrating the
CR-TC.
Hydrating the CR-TC
The CR-TC must be hydrated before initial use by running ASTM
filtered, Type 1 (18 megohm-cm) deionized water through the CR-TC for
10 minutes, while bypassing the columns and suppressor.
Figure 5-62
illustrates the liquid flow path through the CR-TC during the
hydration procedure.
Figure 5-62. CR-TC Plumbing for Hydration