15 vacuum degas assembly does not run – Oritron Dionex Ion Chromatography System ICS-2100 User Manual
Page 109
4 • Troubleshooting
Doc. 065291-01 3/09
97
•
Cell above or below temperature
Contact Dionex for assistance.
•
Damaged piston seal
Replace the piston seal (see
4.15 Vacuum Degas Assembly Does Not Run
•
Degas option not enabled in Chromeleon
1. Open the Chromeleon Server Configuration program. Right-click the
ICS-2100 device in the timebase and select Properties.
2. Select the Options tab (see
Figure 4-3
).
.
Figure 4-3. ICS-2100 Server Configuration Properties: Options
3. Verify that the Degas check box is selected.