Oritron Dionex Ion Chromatography System ICS-2100 User Manual
Page 141
5 • Service
Doc. 065291-01 3/09
129
To calibrate the cell on the ICS-2100 touch screen
DIAGNOSTIC
page:
a.
Go to the
DIAGNOSTIC
page and touch
Electric Conductivity Cell
.
b. On the
ELECTRIC CONDUCTIVITY CALIBRATION
page (see
Figure 5-20
),
touch
Run Offset Cal
.
c.
When the offset calibration is complete, touch
Run Slope Cal
.
d. Touch the
Cell Constant
field and enter the value printed on the front of
the cell.
e.
To ensure that the new calibration values are recorded in the ICS-2100
memory, do not turn off the ICS-2100 power for at least 1 minute.
f.
Record the new calibration values in the Chromeleon Wellness database.
To do this, open the Wellness Panel (see
) and click Upload
under Update Wellness Database.
Figure 5-20. Electric Conductivity Calibration Page