A.9 injection valve, A.10 auxiliary valve (optional), A.11 vacuum degas assembly – Oritron Dionex Ion Chromatography System ICS-2100 User Manual
Page 214: A.12 column heater, A.13 auxiliary power supply (optional), Injection valve, A.9 injection valve a.10 auxiliary valve, Optional) a.11 vacuum degas assembly
ICS-2100 Ion Chromatography System
202
Doc. 065291-01 3/09
A.9
Injection Valve
A.10 Auxiliary Valve
(
Optional)
A.11 Vacuum Degas Assembly
A.12 Column Heater
A.13 Auxiliary Power Supply (Optional)
Injection Valve
6-port, 2-position Rheodyne valve with PEEK wetted components;
electrically-activated
Auxiliary Valve
6-port or 10-port, 2-position Rheodyne valve with PEEK wetted
components; electrically-activated
Channel
Single-channel degas channel with degas membranes
Pump
Dual-stage diaphragm vacuum pump
Materials
Wetted materials, PEEK, Teflon
Operating
Temperature
30 to 60 °C (86 to 140 °F); minimum 5 °C above ambient
temperature. Settable range is equal to working range.
Temperature
Stability
±0.5 °C at sensor
Temperature
Accuracy
±0.5 °C at sensor, at 40 °C
Maximum
Column Lengths
250 mm analytical column with 50 mm guard column
Constant Current
200 mA maximum
Voltage
35 V maximum