Oritron Dionex Ion Chromatography System ICS-2100 User Manual
Page 189
5 • Service
Doc. 065291-01 3/09
177
4. Connect the pump outlet to the mixer inlet (see
Figure 5-69
).
5. Disconnect the mixer outlet line (labeled
TO INJ VALVE PORT P
) from
port
P
(
2
) on the injection valve.
6. Direct the mixer outlet line to a temporary waste container, such as a
beaker.
7. From the ICS-2100 Control panel in Chromeleon or the ICS-2100
touch screen
HOME
page, set the pump flow rate to 5.00 mL/min.
8. Verify that the suppressor and the auxiliary electrolytic device (if
installed) are off.
9. Verify that the
Automatically turn on with pump
option is not
selected on the touch screen
SUPPRESSOR
Disabling this option lets you turn on the pump, without
automatically turning on the suppressor.
10. Turn on the pump.
Figure 5-69. Pump Outlet Connected to Mixer Inlet for Mixer Hydration
To avoid damaging the suppressor and auxiliary electrolytic device (if
installed), always turn them off before filling the mixer. The pump flow
is on during filling; however, no flow reaches the suppressor or
auxiliary electrolytic device.
Mixer Inlet
Pump Outlet