Ics-2100 ion chromatography system – Oritron Dionex Ion Chromatography System ICS-2100 User Manual
Page 262
ICS-2100 Ion Chromatography System
250
Doc. 065291-01 3/09
Timed Event Notes:
•
At time INIT, the injection valve is set to Load.
•
At time 0.00, the injection valve is set to Inject, an autozero is performed
(TTL2=1), the data acquisition mark is sent (RLY1=1), and the pump is
turned on (RLY2=1).
•
At time 0.5, TTL2 and RLY1 are switched back to 0 to be ready for the next
run. RLY2 remains at 1 to keep the pump on.
•
TTL1 does not need to be set in the timed events because the Load and Inject
commands control the ICS-2100 injection valve automatically (see
“Enable
AS Control of the Injection Valve” on page 248
).
To run the above method, set up an AS schedule. The example in
Figure C-11
includes a four-point calibration (Lines 1–4) followed by two samples (Lines 5
and 6).
Figure C-11. Example AS Schedule for Stand-Alone Operation
Help Prompt
SAVE TO
RUN
SCHEDULE
25.0
5.0
1
1
1
4
3
5
5
1
6
6
1
1
MISSING VIAL ACTION:
INJ/
INJ
VIAL
METHOD
VOL(uL)
VIAL#
LINE
END
START
EDIT
SKIP
1
1
1
1-4
5
6
-
-
-
-
25.0