Oritron Dionex Ion Chromatography System ICS-2100 User Manual
Page 260
ICS-2100 Ion Chromatography System
248
Doc. 065291-01 3/09
Assign ICS-2100 Input Functions
Go to the ICS-2100
INPUT/OUTPUT
touch screen page and verify that the
Inject
function is assigned to TTL input 1, that Autozero is assigned to TTL input 2, and
that the
TTL Input Control Type
is
Normal Edge
(see
Figure C-7
).
Enable AS Control of the Injection Valve
1. On the AS front panel, press
Menu
and
0
to go to the
TIME FUNCTION OUT
screen. In the
LOAD/INJECT VALVE: TTL 1
field, select
ENABLED
.
2. Turn the AS power off and then back on.
You can now issue commands from the AS to switch the ICS-2100 injection
valve position.
Figure C-7. Input/Output Page
Figure C-8. AS Time Function Out Screen
Help Prompt
TIME FUNCTION OUT
TTL
1
2
LOAD/INJECT VALVE:
LOAD/INJECT VALVE:
ENABLED
DISABLED
STATE