A.3 environmental, A.4 front panel, A.5 analytical pump and fluidics – Oritron Dionex Ion Chromatography System ICS-2100 User Manual
Page 208: Environmental, Front panel, Analytical pump and fluidics, Refer to, Section a.3
ICS-2100 Ion Chromatography System
196
Doc. 065291-01 3/09
A.3
Environmental
A.4
Front Panel
A.5
Analytical Pump and Fluidics
Operating
Temperature
4 to 40 °C (40 to 104 °F); cold room-compatible (4 °C) as long
as system power remains on
Operating
Humidity
5% to 95% relative humidity, noncondensing
Operating Pressure
35 MPa (5000 psi) maximum liquid path (tubing, valve,
columns, etc.)
Power LED
On when power is present; off when no power is present
LCD Touch
Screen
¼ VGA (320 x 240 pixels) liquid crystal display with touch screen,
which provides:
•
Local control of ICS-2100 components: pump, detector,
suppressor, injection valve, eluent generator, CR-TC, column
heater, vacuum degas, TTLs, and relays
•
Display of conductivity data plot and operating status
•
Local control of calibration and diagnostic functions
Type
Serial dual-reciprocating pistons, microprocessor-controlled
constant stroke, variable speed
Construction
Chemically inert, metal-free PEEK pump heads and flow paths
compatible with aqueous eluents of pH 0 to 14 and reversed-phase
solvents
Operating
Pressure
0 to 35 MPa (0 to 5000 psi)
Flow Rate
Range
0.00 to 5.00 mL/min in 0.01 mL/min increments without changing
pump heads
Flow Precision
<0.1% typical