Oritron Dionex Ion Chromatography System ICS-2100 User Manual
Page 198
ICS-2100 Ion Chromatography System
186
Doc. 065291-01 3/09
4. Remove the
EGC
VENT
line (see
Figure 5-73
) from
the EluGen cartridge
by unscrewing the
luer lock from the
luer adapter.
If necessary, use a
wrench to hold the
luer adapter in place
while unscrewing
the vent line.
Figure 5-73. EluGen Cartridge Vent Line and Luer Fitting
5. Install the luer cap (P/N 053981), provided with the EluGen cartridge,
onto the luer adapter. This prevents leaks from the vent opening when
you turn over the EluGen cartridge (see
Step 8
).
6. If an EPM is installed, lift up the bracket with the EPM attached and
remove it from the EGC holder.
7. Lift the EluGen cartridge straight up and out of the EGC holder.
8. Turn the cartridge upside down and set it in the EluGen cartridge
service area on top of the ICS-2100 (see
Figure 5-36
).
Figure 5-74. EluGen Cartridge in Service Area
Luer Lock
Luer Adapter
EGC Vent
Line