Oritron Dionex Ion Chromatography System ICS-2100 User Manual
Page 201
5 • Service
Doc. 065291-01 3/09
189
5. Check that none of the
lines are caught under the
CR-TC and then push the
CR-TC onto the metal
stud inside the EGC
holder. Push down until
the CR-TC snaps in place
(see
Figure 5-78
).
Figure 5-78. CR-TC Installed in Holder
6. Feed the CR-TC cable
through the slot on the
side of the EGC holder
(see
Figure 5-79
).
Figure 5-79. CR-TC Cable
7. Turn the EGC holder right side up and set it on top of the ICS-2100.
8. Direct the
WASTE, GAS SEPARATOR
line from the top of the ICS-
2100 to the rear panel. Snap the line onto one of the tubing clips on
the rear panel. Connect the line to the gas separator waste tube (see
Section B.10.1
).
CR-TC
Cable