7 erratic flow/pressure reading, Erratic flow/pressure reading, Open (see – Oritron Dionex Ion Chromatography System ICS-2100 User Manual
Page 103: Figure 4-2
4 • Troubleshooting
Doc. 065291-01 3/09
91
•
Flow rate is set to 0
Set the flow rate from the ICS-2100 Control panel in Chromeleon or from the
touch screen
HOME
page.
•
Eluent valve is closed
Open the eluent valve from the ICS-2100 Control panel in Chromeleon or
from the touch screen
PUMP
page.
•
Pump not primed
•
Broken pump piston
Replace the piston (see
•
Eluent generator degas tubing assembly ruptured
Replace the EluGen cartridge holder (see
4.7
Erratic Flow/Pressure Reading
•
Pump needs priming
Figure 4-2. Waste and Priming Valves
Waste Valve
Priming Valve