Orbita, Head orbita – Siemens syngo CT 2005C User Manual
Page 84

84
Head
Orbita
Indications:
Spiral mode for orbital studies, e.g., fracture.
A range of 2 cm will be covered in 12 sec.
Orbita
2
nd
reconstr.
kV
130
Effective mAs
100
Rotation time
1.0 sec
Slice collimation
1.0 mm
Slice width
3.0 mm
1.25 mm
Feed/Rotation
2.0 mm
Pitch Factor
1.0
Increment
3.0 mm
0.8 mm
Kernel
H70s
H70s
CTDI
Vol
22.77 mGy
Effective dose
Male:
0.10 mSv
Female:
0.11 mSv
Contrast medium IV injection
Start delay
60 sec
Flow rate
2 ml/sec
Volume
50 – 60 ml
C2-025.630.01.01.02_APPLICATIONGUIDE_SPIRIT.book Page 84 Friday, April 8, 2005 9:55 AM