179
A range for the Thorax of 25 cm will be covered in
15 sec., a range for the AbdPelvis of 20 cm will be cov-
ered in 12 sec.
• For long range scanning, please pay attention to the
mark of scannable range on the table mattress while
positioning the patient.
• In some cases, it might be advisable to position the
patient feet first so that there will be more space for
the intensive care equipment around.
Thorax
AbdPelvis
kV
130
130
Effective mAs
40
60
Rotation time
1.0 sec
1.0 sec
Slice collimation
5.0 mm
5.0 mm
Slice width
8.0 mm
8.0 mm
Feed/Rotation
20.0 mm
20.0 mm
Pitch Factor
2.0
2.0
Increment
8.0 mm
8.0 mm
Kernel
B41s
B41s
CTDI
Vol
4.31 mGy
6.47 mGy
Effective dose
Male:
1.82 mSv
Female:
2.45 mSv
Male:
2.59 mSv
Female:
3.31 mSv