A range for the Thorax of 25 cm will be covered in
15 sec., a range for the AbdPelvis of 20 cm will be cov-
ered in 12 sec.

• For long range scanning, please pay attention to the

mark of scannable range on the table mattress while
positioning the patient.

• In some cases, it might be advisable to position the

patient feet first so that there will be more space for
the intensive care equipment around.

Thorax

AbdPelvis

kV

130

130

Effective mAs

40

60

Rotation time

1.0 sec

1.0 sec

Slice collimation

5.0 mm

5.0 mm

Slice width

8.0 mm

8.0 mm

Feed/Rotation

20.0 mm

20.0 mm

Pitch Factor

2.0

2.0

Increment

8.0 mm

8.0 mm

Kernel

B41s

B41s

CTDI

Vol

4.31 mGy

6.47 mGy

Effective dose

Male:

1.82 mSv

Female:

2.45 mSv

Male:

2.59 mSv

Female:

3.31 mSv

C2-025.630.01.01.02_APPLICATIONGUIDE_SPIRIT.book Page 179 Friday, April 8, 2005 9:55 AM