Thorax – Siemens syngo CT 2005C User Manual
Page 105

Thorax
105
Hints
• You could repeat the same protocol by simply click-
ing the chronicle with the right mouse button for
“repeat“, e.g., when both non-contrast and contrast
studies are required.
• For lung cancer evaluation, this protocol can be com-
bined with the protocol “Neck Routine”.
ThorFast
2
nd
reconstr.
kV
130
Effective mAs
60
Rotation time
0.8 sec
Slice collimation
5.0 mm
Slice width
8.0 mm
8.0 mm
Feed/Rotation
20.0 mm
Pitch Factor
2.0
Increment
8.0 mm
8.0 mm
Kernel
B41s
B70s
CTDI
Vol
6.47 mGy
Effective dose
Male:
3.38 mSv
Female:
4.39 mSv
Contrast medium IV injection
Start delay
25 – 30 sec.
Flow rate
2.5 ml/sec.
Total amount
80 ml
C2-025.630.01.01.02_APPLICATIONGUIDE_SPIRIT.book Page 105 Friday, April 8, 2005 9:55 AM