Thorax
103
Hints
• You could repeat the same protocol by simply click-
ing the chronicle with the right mouse button for
“repeat“, e.g., when both non-contrast and contrast
studies are required.
• For lung cancer evaluation, this protocol can be com-
bined with the protocol “Neck Routine”.
ThorRoutine
2
nd
reconstruction
kV
130
Effective mAs
60
Rotation time
0.8/1.0 sec
Slice
collimation
4.0 mm
Slice width
5.0 mm
5.0 mm
Feed/Rotation
14.4 mm
Pitch Factor
1.8
Increment
5.0 mm
5.0 mm
Kernel
B41s
B70s
CTDI
Vol
6.47 mGy
Effective dose
Male:
3.28 mSv
Female:
4.23 mSv
Contrast medium IV injection
Start delay
25 – 30 sec.
Flow rate
2.5 ml/sec.
Total amount
80 ml