Spine – Siemens syngo CT 2005C User Manual
Page 143

Spine
143
L3-4
L4-5
L5-S1
kV
130
130
130
Effective mAs
220
220
235
Rotation time
1.5 sec
1.5 sec
1.5 sec
Slice
collimation
1.5 mm
1.5 mm
1.5 mm
Slice width
3.0 mm
3.0 mm
3.0 mm
Feed/Scan
3.0 mm
3.0 mm
3.0 mm
Kernel
B31s
B31s
B31s
CTDI
Vol
23.72
mGy
23.72
mGy
25.33
mGy
Effective dose
Male:
0.58 mSv
Female:
0.92 mSv
Male:
0.63 mSv
Female:
1.21 mSv
Male:
0.64 mSv
Female:
2.35 mSv
C2-025.630.01.01.02_APPLICATIONGUIDE_SPIRIT.book Page 143 Friday, April 8, 2005 9:55 AM