Dental, Head dental – Siemens syngo CT 2005C User Manual
Page 86
86
Head
Dental
This is the scan protocol for the syngo Dental CT appli-
cation package. It is used for evaluation and reformat-
ting of the upper and lower jaws.
It enables the display and measurement of the bone
structures of the upper and lower jaw as the basis for
planning in oral surgery.
Indications:
Spiral mode for dental studies.
A range of 5 cm will be covered in 27 sec.
Load the study into the application “syngo Dental CT”.
For further information, please refer to the chapter
"syngo Dental CT".
Dental
kV
130
Effective mAs
45
Rotation time
1.0 sec
Slice collimation
1.0 mm
Slice width
1.25 mm
Feed/Rotation
2.0 mm
Pitch Factor
1.0
Increment
0.5 mm
Kernel
H60s
CTDI
Vol
10.25 mGy
Effective dose
Male: 0.13 mSv
Female: 0.16 mSv
C2-025.630.01.01.02_APPLICATIONGUIDE_SPIRIT.book Page 86 Friday, April 8, 2005 9:55 AM