beautypg.com

Syngo volume evaluation – Siemens syngo CT 2005C User Manual

Page 302

background image

302

syngo Volume Evaluation

Restrictions for Linking and Unlinking

There are some restrictions for linking and unlinking:
– Linking can be done on user-defined ROIs as well

as on interpolated ROIs. If you link an interpolated
ROI (dotted line), it becomes a user-confirmed ROI
(solid line).

– You cannot unlink interpolated ROIs. The unlinking

only works with user-confirmed ROIs.

– If you modify the volume (for example, move a

ROI) before linking an ROI to the already selected
ROI, the marking is reset.

• Evaluation
– Bounding voxels, where no voxel lies within the

entered range of HU values, stops region growing.

Check the depth of the volume in the coronal and
sagittal view. To complete region growing inside the
entire volume, set multiple seed points if necessary.

– You can edit the thresholds even after the volume

has been evaluated.

– You can modify or delete the last evaluated volume.

C2-025.630.01.01.02_APPLICATIONGUIDE_SPIRIT.book Page 302 Friday, April 8, 2005 9:55 AM