beautypg.com

Headtrauma, Specials headtrauma – Siemens syngo CT 2005C User Manual

Page 180

background image

180

Specials

HeadTrauma

A spiral mode for emergency head studies with a max.
FoV of 500 mm. A typical scan range of 40/80 cm will
be covered in 23/27 sec.

Base

Cerebrum

kV

130

130

Effective mAs

110

110

Rotation time

1.5 sec

1.0 sec

Slice collimation

1.5 mm

2.5 mm

Slice width

3.0 mm

8.0 mm

Feed/Rotation

3.0 mm

5.0 mm

Pitch Factor

1.0

1.0

Increment

3.0 mm

8.0 mm

Kernel

H31s

H31s

CTDI

Vol

25.05 mGy

25.05 mGy

Effective dose

Male:

0.37 mSv

Female:
0.51 mSv

Male:

0.70 mSv

Female:
0.86 mSv

C2-025.630.01.01.02_APPLICATIONGUIDE_SPIRIT.book Page 180 Friday, April 8, 2005 9:55 AM