C-spineseq, Spine c-spineseq – Siemens syngo CT 2005C User Manual
Page 138
138
Spine
C-SpineSeq
Indications:
Sequential mode for cervical spine studies, e.g., pro-
lapse, degenerative changes, trauma, tumors, etc.
with a typical length for each range of 1.2 cm.
This protocol contains five ranges:
C2-C3, C3-C4, C4-C5, C5-C6, C6-C7
C2-3
C3-4
C4-5
kV
130
130
130
Effective mAs
170
170
170
Rotation time
1.5 sec
1.5 sec
1.5 sec
Slice
collimation
1.0 mm
1.0 mm
1.0 mm
Slice width
2.0 mm
2.0 mm
2.0 mm
Feed/Scan
2.0 mm
2.0 mm
2.0 mm
Kernel
B50s
B50s
B50s
CTDI
Vol
18.33
mGy
18.33
mGy
18.33
mGy
Effective dose
Male:
0.18 mSv
Female:
0.14 mSv
Male:
0.39 mSv
Female:
0.21 mSv
Male:
0.66 mSv
Female:
0.45 mSv
C2-025.630.01.01.02_APPLICATIONGUIDE_SPIRIT.book Page 138 Friday, April 8, 2005 9:55 AM