Spine – Siemens syngo CT 2005C User Manual
Page 139

Spine
139
Hints
• Dental artifacts can be reduced by tilting the gantry.
• For MPRs and 3D post-processing, use spiral mode
with thinner slice and reconstruct images with 50%
overlap in image reconstruction. Define a long range
to cover the levels of interest in one block.
C5-6
C6-7
kV
130
130
Effective mAs
170
190
Rotation time
1.5 sec
1.5 sec
Slice collimation
1.0 mm
1.0 mm
Slice width
2.0 mm
2.0 mm
Feed/Scan
2.0 mm
2.0 mm
Kernel
B50s
B50s
CTDI
Vol
18.33 mGy
20.48 mGy
Effective dose
Male:
0.66 mSv
Female:
0.73 mSv
Male:
0.30 mSv
Female:
0.50 mSv
C2-025.630.01.01.02_APPLICATIONGUIDE_SPIRIT.book Page 139 Friday, April 8, 2005 9:55 AM