Indications:
For soft tissue spiral studies in the cervical region, e.g.,
tumors, lymphoma, abscesses etc.
A typical range of 20 cm will be covered in 42 sec.
Neck
kV
130
Effective mAs
70
Rotation time
1.0 sec
Slice collimation
2.5 mm
Slice width
5.0 mm
Feed/Rotation
5.0 mm
Pitch Factor
1.0
Increment
5.0 mm
Kernel
B50s
CTDI
Vol
7.55 mGy
Effective dose
Male: 1.79 mSv
Female: 1.87 mSv
Contrast medium IV injection
Start delay
45 sec.
Flow rate
2.0 ml/sec.
Total amount
120 ml
C2-025.630.01.01.02_APPLICATIONGUIDE_SPIRIT.book Page 92 Friday, April 8, 2005 9:55 AM