beautypg.com

Carotidangio08s, Children carotidangio08s – Siemens syngo CT 2005C User Manual

Page 232

background image

232

Children

CarotidAngio08s

Indications:

Spiral mode for the carotid arteries, e.g., carotid steno-
sis or occlusion, vascular abnormalities of the carotids
or vertebral arteries, etc.

A range of 10 cm will be covered in 22 sec.

* The conversion factor for a 7-year-old child, and a
scan range of 100 mm was used.

CarotidAngio

2

nd

reconstr.

kV

80

Effective mAs

50

Rotation time

0.8 sec

Slice

collimation

1.0 mm

Slice width

3.0 mm

1.25 mm

Feed/Rotation

4.0 mm

Pitch Factor

2.0

Increment

3.0 mm

0.8 mm

Kernel

B41s

B41s

CTDI

Vol

1.38 mGy

Effective dose

Male:

0.50 mSv*

Female:

0.50 mSv*

Contrast medium IV injection
Start delay

exam dependent

Flow rate

dependent upon needle size/

Access site

Total amount

1 – 2 ml per kg of body weight

C2-025.630.01.01.02_APPLICATIONGUIDE_SPIRIT.book Page 232 Friday, April 8, 2005 9:55 AM