Carotidangio08s, Children carotidangio08s – Siemens syngo CT 2005C User Manual
Page 232
232
Children
CarotidAngio08s
Indications:
Spiral mode for the carotid arteries, e.g., carotid steno-
sis or occlusion, vascular abnormalities of the carotids
or vertebral arteries, etc.
A range of 10 cm will be covered in 22 sec.
* The conversion factor for a 7-year-old child, and a
scan range of 100 mm was used.
CarotidAngio
2
nd
reconstr.
kV
80
Effective mAs
50
Rotation time
0.8 sec
Slice
collimation
1.0 mm
Slice width
3.0 mm
1.25 mm
Feed/Rotation
4.0 mm
Pitch Factor
2.0
Increment
3.0 mm
0.8 mm
Kernel
B41s
B41s
CTDI
Vol
1.38 mGy
Effective dose
Male:
0.50 mSv*
Female:
0.50 mSv*
Contrast medium IV injection
Start delay
exam dependent
Flow rate
dependent upon needle size/
Access site
Total amount
1 – 2 ml per kg of body weight