Syngo osteo ct – Siemens syngo CT 2005C User Manual
Page 277
syngo Osteo CT
277
• Position the cut line of scanning through the middle
of the vertebra, i.e. bi-sector between the angle of
the upper and lower end plate.
• The phantom must be included in the FoV of the
images for evaluation.
• It is recommended to end the exam first, and then
start the syngo Osteo CT evaluation.
• Do not use the calibration phantom from other CT
scanners, as your system is calibrated to a particular
phantom.
Topographic
Phantom inside
the FoV
C2-025.630.01.01.02_APPLICATIONGUIDE_SPIRIT.book Page 277 Friday, April 8, 2005 9:55 AM