Shoulder

97

Hints

• Use raw data to review a target region if necessary.

• For image reconstruction of soft tissue, use kernel

B31s and a slice width of 5.0 mm.

• Coronal and sagittal 2D planar reconstructions are

important for evaluation of the joint space & bursa
sacs in CT arthograms.

• 3D renderings are helpful for complex fractures &

dislocations.

Shoulder

2

nd

reconstruction

kV

130

Effective mAs

70

Rotation time

1.0 sec

Slice
collimation

1.5 mm

Slice width

3.0 mm

2.0 mm

Feed/Rotation

3.0 mm

Pitch Factor

1.0

Increment

3.0 mm

1.5 mm

Kernel

B60s

CTDI

Vol

7.55 mGy

Effective dose

Male:
0.94 mSv

Female:
1.13 mSv