312
syngo Dynamic Evaluation
– Peak Enhancement image: Parameter image where
each pixel indicates the enhancement at that pixel
location.
CT value of a pixel = maximum value found at that
pixel location (pixel value of the MIP image) – CT
value of the baseline image at that pixel location.
With a Peak Enhancement image, for example, you
can easily detect structures that were not completely
filled with contrast.
Hint: You can activate or deactivate the calculation of
the Time to Peak image and the Peak Enhancement
image in the configuration.
C2-025.630.01.01.02_APPLICATIONGUIDE_SPIRIT.book Page 312 Friday, April 8, 2005 9:55 AM