Children
209
* The conversion factor for a 7-year-old child, and a
scan range of 150 mm was used.
Hints
• Children with a body weight of more than 55 kg
should be examined with an adult protocol.
• The first and second recon jobs are defined for visu-
alization of the mediastinum and the lungs, respec-
tively.
• Please change the mAs value according to the
weight of the child:
45 – 54 kg 37 mAs.
ThorRoutine
2
nd
reconstruction
kV
130
Effective mAs
26
Rotation time
1.0 sec
Slice
collimation
2.5 mm
Slice width
5.0 mm
5.0 mm
Feed/Rotation
10.0 mm
Pitch Factor
2.0
Increment
5.0 mm
5.0 mm
Kernel
B41s
B60s
CTDI
Vol
2.80 mGy
Effective dose
Male: 1.46 mSv*
Female: 1.72 mSv*
Contrast medium IV injection
Start delay
exam dependent
Flow rate
dependent upon needle size/
Access site
Total amount
1 – 2 ml per kg of body weight