Children – Siemens syngo CT 2005C User Manual
Page 207

Children
207
* The conversion factor for an 8-week-old child, and a
scan range of 150 mm was used.
Hints
• Use this protocol for children below 35 kg.
• The first and second recon jobs are defined for visu-
alization of the mediastinum and the lungs, respec-
tively.
• Please change the mAs value according to the
weight of the child:
25 – 34 kg 60 mAs.
ThorRoutine
2
nd
reconstruction
kV
80
Effective mAs
41
Rotation time
1.0 sec
Slice
collimation
2.5 mm
Slice width
5.0 mm
5.0 mm
Feed/Rotation
10.0 mm
Pitch Factor
2.0
Increment
5.0 mm
5.0 mm
Kernel
B41s
B60s
CTDI
Vol
1.13 mGy
Effective dose
Male: 1.55 mSv*
Female: 1.85 mSv*
Contrast medium IV injection
Start delay
exam dependent
Flow rate
dependent upon needle size/
Access site
Total amount
1 – 2 ml per kg of body weight
C2-025.630.01.01.02_APPLICATIONGUIDE_SPIRIT.book Page 207 Friday, April 8, 2005 9:55 AM