Abdomen – Siemens syngo CT 2005C User Manual
Page 119

Abdomen
119
* CARE Bolus may be used to optimize the bolus tim-
ing. Set the ROI for monitoring scan in the abdominal
aorta at the beginning of the scan range with trigger-
ing threshold of 120 HU, or use manual triggering.
Non
Contrast
Arterial
Phase
Venous
Phase
kV
130
130
130
Effective mAs
80
80
80
Rotation time
1.0 sec
0.8/1.0
sec
1.0 sec
Slice
collimation
5.0 mm
5.0 mm
5.0 mm
Slice width
8.0 mm
6.0 mm
6.0 mm
Feed/Rotation
15.0 mm
14.0/
15.0 mm
15.0 mm
Pitch Factor
1.5
1.4/1.5
1.5
Increment
8.0 mm
6.0 mm
6.0 mm
Kernel
B41s
B41s
B41s
CTDI
Vol
8.62 mGy
8.62 mGy
8.62 mGy
Effective dose
Male:
3.34 mSv
Female:
4.19 mSv
Male:
3.34 mSv
Female:
4.19 mSv
Male:
3.34 mSv
Female:
4.19 mSv
Contrast medium IV injection
Start delay
20 – 25 sec.* (arterial phase)
50 – 75 sec.* (venous phase)
Flow rate
3.0 ml/sec.
Total amount
100 ml