Children

ThoraxHRSeq_Child

Indications:

Sequence mode for High Resolution lung studies, e.g.,
interstitial changes of the lung parenchyma, etc.

A range of 20.1 cm is predefined.

* The conversion factor for a 7-year-old child, and a
scan range of 201 mm was used.

ThorHRSeq

kV

130

Effective mAs

70

Rotation time

1.0 sec

Slice collimation

1.0 mm

Slice width

1.0 mm

Feed/Scan

10.0 mm

Kernel

U90s

CTDI

Vol

1.51 mGy

Effective dose

Male: 1.02 mSv*

Female: 1.21 mSv*