Thoraxhrseq_child, Children thoraxhrseq_child – Siemens syngo CT 2005C User Manual
Page 212
212
Children
ThoraxHRSeq_Child
Indications:
Sequence mode for High Resolution lung studies, e.g.,
interstitial changes of the lung parenchyma, etc.
A range of 20.1 cm is predefined.
* The conversion factor for a 7-year-old child, and a
scan range of 201 mm was used.
ThorHRSeq
kV
130
Effective mAs
70
Rotation time
1.0 sec
Slice collimation
1.0 mm
Slice width
1.0 mm
Feed/Scan
10.0 mm
Kernel
U90s
CTDI
Vol
1.51 mGy
Effective dose
Male: 1.02 mSv*
Female: 1.21 mSv*
C2-025.630.01.01.02_APPLICATIONGUIDE_SPIRIT.book Page 212 Friday, April 8, 2005 9:55 AM