Vascular – Siemens syngo CT 2005C User Manual
Page 171
Vascular
171
Hints
• CARE Bolus may be used to optimize the bolus tim-
ing.
• Set the ROI for monitoring scan in the pulmonary
trunk with triggering threshold of 120 HU, or use
manual triggering.
• You could repeat the same protocol by simply click-
ing the chronicle with the right mouse button for
"repeat".
Contrast medium IV injection
Start delay
6 – 10 sec.
Flow rate
4 ml/sec.
Total amount
80 – 100 ml