Children – Siemens syngo CT 2005C User Manual

Page 203

background image

Children

203

Hints

• Children, who are more than 6 years old, should be

scanned with an adult protocol.

• In order to optimize image quality versus radiation

dose, scans are provided within a maximum scan
field of 300 mm with respect to the iso-center. No
recon job with a field of view exceeding those limits
will be possible. Therefore, patient positioning has to
be performed accurately to ensure a centered loca-
tion of the skull.

• Please change the mAs value according to the age of

the child:

3 – 6 years 100 mAs.

Contrast medium IV injection
Start delay

exam dependent

Flow rate

dependent upon needle size/

Access site

Total amount

1 – 2 ml per kg of body weight