beautypg.com

Test bolus using care bolus, Workflow information – Siemens syngo CT 2005C User Manual

Page 50

background image

50

Workflow Information

Test Bolus using CARE Bolus

You can use the CARE Bolus option as a “Test Bolus“.

How to do it

1.Insert a Bolus tracking via the right mouse button

submenu prior to the spiral.

2.Insert “contrast“ from the right mouse button con-

text menu prior to the monitoring scans.

Hint: By inserting “contrast“ you are interrupting the
Auto range function, and therefore an automatic
start of the spiral is not possible!

3.Start with the Topogram.

4.Position the premonitoring scan and the spiral

range.

5.Perform the premonitoring scan, position and

accept the ROI.

6.Start monitoring scans and a short amount of con-

trast (20 ml/2.5 ml/sec.).

Hint: With starting the spiral the system is switching
to the Trigger subtask card. The trigger line is not
shown at this stage.

7.Now you can read the proper delay from the Trigger

subtask card.

8. Insert the delay in the Routine subtask card and load

the spiral.

9.Start spiral and injector with the full amount of con-

trast.

C2-025.630.01.01.02_APPLICATIONGUIDE_SPIRIT.book Page 50 Friday, April 8, 2005 9:55 AM