Neonatebody, Children neonatebody – Siemens syngo CT 2005C User Manual
Page 238
238
Children
NeonateBody
Indications:
Spiral mode for routine neonate body studies, e.g.,
tumors, abnormalities, malformations, abscesses, etc.
A range of 15 cm will be covered in 17 sec.
* The conversion factor for a 7-year-old child, and a
scan range of 150 mm was used.
NeonateBody
kV
80
Effective mAs
41
Rotation time
1.0 sec
Slice collimation
2.5 mm
Slice width
5.0 mm
Feed/Rotation
10.0 mm
Pitch Factor
2.0
Increment
5.0 mm
Kernel
B41s
CTDI
Vol
1.13
Effective dose
Male: 1.59 mSv*
Female: 2.01 mSv*
Contrast medium IV injection
Start delay
exam dependent
Flow rate
dependent upon needle size/
Access site
Total amount
1 – 2 ml per kg of body weight
C2-025.630.01.01.02_APPLICATIONGUIDE_SPIRIT.book Page 238 Friday, April 8, 2005 9:55 AM