Extrhr_baby, Children extrhr_baby – Siemens syngo CT 2005C User Manual
Page 222
222
Children
ExtrHR_Baby
Indications:
Spiral mode for high resolution extremity studies, e.g.,
post-traumatic changes, tumors, malformations, etc.
A range of 10 cm will be covered in 27 sec.
* The conversion factor for an 8-week-old child, and a
scan range of the knee of 100 mm was used.
ExtrHR
kV
80
Effective mAs
50
Rotation time
1.0 sec
Slice collimation
1.0 mm
Slice width
1.25 mm
Feed/Rotation
4.0 mm
Pitch Factor
2.0
Increment
0.8 mm
Kernel
B70s
CTDI
Vol
1.38 mGy
Effective dose
Male: < 0.01 mSv*
Female: < 0.01 mSv*
C2-025.630.01.01.02_APPLICATIONGUIDE_SPIRIT.book Page 222 Friday, April 8, 2005 9:55 AM