Care dose, How does care dose work, Dose information care dose – Siemens syngo CT 2005C User Manual
Page 32
32
Dose Information
CARE Dose
CARE Dose is a clinical application package that pro-
vides real-time tube current modulation for Spiral and
Sequential Scanning.
CARE Dose reduces patient dose significantly, espe-
cially in the regions of shoulder and pelvis. It decreases
tube load, which extends the capacity for volume scan-
ning with thinner slices, larger volumes or Multi-phase
studies.
It can also improve image quality by increasing mA,
thus reducing image noise on the lateral views.
How does CARE Dose work?
It reduces the mA for low attenuation views up to 90%
and keeps the nominal higher mA for high attenuation
views, e. g. in the lateral projection. This is done “on-
the-fly”, i.e. the scanner adapts the mA in real-time,
according to the patient’s attenuation profile.
C2-025.630.01.01.02_APPLICATIONGUIDE_SPIRIT.book Page 32 Friday, April 8, 2005 9:55 AM