beautypg.com

Pelvis – Siemens syngo CT 2005C User Manual

Page 129

background image

Pelvis

129

Hints

• In case of 3D study only, images should be recon-

structed with at least 50% overlap in image recon-
struction and kernel B20s.

• If only one side is examined, it is advisable to enter

the side in the comment line on the Routine card.

Hip

2

nd

reconstr.

kV

130

Effective mAs

70

Rotation time

1.0 sec

Slice collimation

1.5 mm

Slice width

3.0 mm

2.0 mm

Feed/Rotation

3.0 mm

Pitch Factor

1.0

Increment

3.0 mm

1.5 mm

Kernel

B70s

B70s

CTDI

Vol

7.55 mGy

Effective dose

Male:

1.78 mSv

Female:

1.66 mSv

C2-025.630.01.01.02_APPLICATIONGUIDE_SPIRIT.book Page 129 Friday, April 8, 2005 9:55 AM