Pelvis – Siemens syngo CT 2005C User Manual
Page 129
Pelvis
129
Hints
• In case of 3D study only, images should be recon-
structed with at least 50% overlap in image recon-
struction and kernel B20s.
• If only one side is examined, it is advisable to enter
the side in the comment line on the Routine card.
Hip
2
nd
reconstr.
kV
130
Effective mAs
70
Rotation time
1.0 sec
Slice collimation
1.5 mm
Slice width
3.0 mm
2.0 mm
Feed/Rotation
3.0 mm
Pitch Factor
1.0
Increment
3.0 mm
1.5 mm
Kernel
B70s
B70s
CTDI
Vol
7.55 mGy
Effective dose
Male:
1.78 mSv
Female:
1.66 mSv
C2-025.630.01.01.02_APPLICATIONGUIDE_SPIRIT.book Page 129 Friday, April 8, 2005 9:55 AM