Thorangio/thorangio08s, Vascular thorangio/thorangio08s – Siemens syngo CT 2005C User Manual
Page 168
168
Vascular
ThorAngio/ThorAngio08s
Indications:
Spiral study for location and extent of aneurysms, dis-
section, and ruptures of the thoriac aorta.
A range of 20 cm will be covered in 24/19 sec.
ThorAngio
2
nd
reconstruction
kV
130
Effective mAs
55
Rotation time
0.8/1.0 sec
Slice collimation
2.5 mm
Slice width
5.0 mm
3.0 mm
Feed/Rotation
9.0 mm
Pitch Factor
1.8
Increment
5.0 mm
2.0 mm
Kernel
B31s
B31s
CTDI
Vol
5.93 mGy
Effective dose
Male: 1.91 mSv
Female: 2.67 mSv
C2-025.630.01.01.02_APPLICATIONGUIDE_SPIRIT.book Page 168 Friday, April 8, 2005 9:55 AM