Bodyangio, Children bodyangio – Siemens syngo CT 2005C User Manual
Page 234

234
Children
BodyAngio
Indications:
Spiral mode for abdominal CT Angio studies, e.g., vas-
cular abnormalities, aneurysms, etc.
A range of 10 cm will be covered in 19 sec.
* The conversion factor for a 7-year-old child, and a
scan range of 100 mm was used.
BodyAngio
2
nd
reconstr.
kV
80
Effective mAs
41
Rotation time
1.0 sec
Slice
collimation
1.5 mm
Slice width
3.0 mm
2.0 mm
Feed/Rotation
6.0 mm
Pitch Factor
2.0
Increment
3.0 mm
1.5 mm
Kernel
B41s
B41s
CTDI
Vol
1.13 mGy
Effective dose
Male:
0.73 mSv*
Female:
0.84 mSv*
Contrast medium IV injection
Start delay
exam dependent
Flow rate
dependent upon needle size/
Access site
Total amount
1 – 2 ml per kg of body weight
C2-025.630.01.01.02_APPLICATIONGUIDE_SPIRIT.book Page 234 Friday, April 8, 2005 9:55 AM