Abdomen_child, Children abdomen_child – Siemens syngo CT 2005C User Manual
Page 216

216
Children
Abdomen_Child
Indications:
Spiral mode for routine studies in the region of abdo-
men and pelvis, e.g., tumors, lymphoma, abscesses,
post-traumatic changes, etc.
A range of 30 cm will be covered in 32 sec.
* The conversion factor for a 7-year-old child, and a
scan range of 300 mm was used.
Abdomen
kV
130
Effective mAs
41
Rotation time
1.0 sec
Slice collimation
2.5 mm
Slice width
5.0 mm
Feed/Rotation
10.0 mm
Pitch Factor
2.0
Increment
5.0 mm
Kernel
B41s
CTDI
Vol
4.42 mGy
Effective dose
Male: 4.83 mSv*
Female: 7.40 mSv*
Contrast medium IV injection
Start delay
exam dependent
Flow rate
dependent upon needle size/
Access site
Total amount
1 – 2 ml per kg of body weight
C2-025.630.01.01.02_APPLICATIONGUIDE_SPIRIT.book Page 216 Friday, April 8, 2005 9:55 AM