Status:operation:enable, Status:operation[:event, Status:preset – HP E1418A User Manual
Page 98: Status
STATus:OPERation:ENABle?
STATus:OPERation:ENABle?
returns the value of the mask set for the Operation
Status Register.
Parameters
Returns: 16-bit signed integer (int16).
STATus:OPERation[:EVENt]?
STATus:OPERation[:EVENt]?
returns the value of the contents of the Operation
Status Event Register. Reading the register clears all bits in the register.
Parameters
Returns: 16-bit signed integer (int16).
Comments
•
Bit 0 of this register (decimal 1) is the only defined bit in this register.
•
This bit is set if the
CALibration:CONFigure:RESistance?
,
CALibration
n
:CONFigure:VOLTage?
, or
CALibration
n
:CONFigure:CURRent?
command has been received and the calibration process is not yet complete.
•
*RST Condition: no effect.
•
*CLS
clears the contents of the Operation Status Register.
•
STATus:PRESet
does not affect the contents of the Operation Status Register, but
does disable reporting the Summary bit to the Status Byte Register.
STATus:PRESet
STATus:PRESet
presets all Enable Registers to 0 (zero). The command has no
effect on Condition or Event Registers.
STATus
Chapter 3
HP E1418A SCPI Command Reference 97