beautypg.com

Status:operation:enable, Status:operation[:event, Status:preset – HP E1418A User Manual

Page 98: Status

background image

STATus:OPERation:ENABle?

STATus:OPERation:ENABle?

returns the value of the mask set for the Operation

Status Register.

Parameters

Returns: 16-bit signed integer (int16).

STATus:OPERation[:EVENt]?

STATus:OPERation[:EVENt]?

returns the value of the contents of the Operation

Status Event Register. Reading the register clears all bits in the register.

Parameters

Returns: 16-bit signed integer (int16).

Comments

Bit 0 of this register (decimal 1) is the only defined bit in this register.

This bit is set if the

CALibration:CONFigure:RESistance?

,

CALibration

n

:CONFigure:VOLTage?

, or

CALibration

n

:CONFigure:CURRent?

command has been received and the calibration process is not yet complete.

*RST Condition: no effect.

*CLS

clears the contents of the Operation Status Register.

STATus:PRESet

does not affect the contents of the Operation Status Register, but

does disable reporting the Summary bit to the Status Byte Register.

STATus:PRESet

STATus:PRESet

presets all Enable Registers to 0 (zero). The command has no

effect on Condition or Event Registers.

STATus

Chapter 3

HP E1418A SCPI Command Reference 97

Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com