APPLy

The

APPLy

subsystem enables and outputs a current or voltage on the specified channel.

Subsystem Syntax

APPLy

n

:CURRent <

value

| MIN | MAX | DEF>

:VOLTage <

value

| MIN | MAX | DEF>

APPLy

n

:CURRent

APPLy

n

:CURRent <

value

| MIN | MAX | DEF>

outputs a current on the specified

channel, n.

Parameters

Parameter

Name

Parameter

Type

Range of

Values

Default

APPLy

n

keyword

1 to 8 or 1 to 16

1

<

value

>

numeric (float 32) or

discrete

-0.02 Amps to +0.02 Amps or

MIN | MAX | DEF

0

Comments

In the keyword,

APPLy

n, the valid range of n is:

1 to 8 for

E1418A, E1418A Option 002

1 to 16 for

E1418A Option 001, E1418A Option 003.

Values of n outside these ranges will generate an error.

The

<

value

>

parameter can range from -0.02 to +0.02 Amps.

MIN

sets -0.02 Amps output.

MAX

sets +0.02 Amps output.

DEF

sets 0 Amps output.

This command is affected by the state of the P/J jumper for this channel on the
terminal module. To use the

APPLy

n

:CURRent command:

the P/J jumper must be set to program selectable or, if set to jumper selectable,

the V/I jumper must be set to the current mode.
These jumpers are described on page 31.

Equivalent Commands:

SOURce

n

:FUNCtion:MODE CURR

,

SOURce

n

:CURRent[:LEVel][:IMMediate][:AMPLitude]

,

OUTPut

n

[:STATe] ON

*RST Condition:

If the P/J jumper is set to program selectable:
output mode is set to voltage, 0 volts output, output relay open.

If the P/J jumper is set to jumper selectable:
output mode is set by the V/I jumper, 0 volts or amps, output relay open.

APPLy